Auto-Ranging ATC Conductivity/TDS/Temp/NaCl Meter
The Hanna 50-HI9835 can measure
conductivity, total dissolved solids, NaCl, and temperature, all with one probe.
The temperature of all conductivity and TDS measurements is automatically or
manually compensated for by the meter's ATC circuitry. The temperature
coefficient is user selectable from 0.00 to 6.00 %/°C (for EC and TDS only) and
the TDS factor can be set from 0.40 to 0.80.
The meter is constructed with a waterproof,
rugged housing which allows the versatility required for both laboratory and
field use. Moreover, 50-HI9835 is equipped with a large LCD
that helps the user through operation and calibration with large comprehensive
symbols. Besides displaying the measurements, the LCD simultaneously displays
the temperature. The display gives a low battery warning and switches itself off
after 5 minutes of non-use to save the batteries.
50-HI9835 is supplied complete
with 50-HI76309 Conductivity/TDS probe with 3.3' (1 m) screened cable, AA
size Alkaline Batteries (4 pcs) and instruction manual in a rugged carrying
case.
| Specifications: |
50-HI9835
|
| |
EC
(Autoranging) |
TDS
(Autoranging) |
NaCl |
Temperature |
| Range |
00.0
to 29.99 µS/cm
30.0 to 299.9 µS/cm
300 to 2999 µS/cm
3.00 to 29.99 mS/cm
30.0 to 200.0 mS/cm
Up to 500.0 mS/cm actual (*) Conductivity |
0.00
to 14.99 ppm
15.0 to 149.9 ppm
150 to 1499 ppm
1.50 to 14.99 g/L
15.0 to 100.0 g/L
Up to 400.0 g/L actual (*) TDS
(with 0.80 factor) |
0.0
to 400.0 % |
0.0
to 60.0°C |
| Resolution |
0.01
µS/cm (from 0.00 to 29.99 µS/cm)
0.1 µS/cm (from 30.0 to 299.9 µS/cm)
1 µS/cm (from 300 to 2999 µS/cm)
0.01 mS/cm (from 3.00 to 29.99 mS/cm)
0.1 mS/cm (over 30.0 mS/cm) |
0.01
ppm (from 0.00 to 14.99 ppm)
0.1 ppm (from 15.0 to 149.9 ppm)
1 ppm (from 150 to 1499 ppm)
0.01 g/L (from 1.50 to 14.99 g/L)
0.1 g/L (over 15.0 g/L) |
0.1
% |
0.1°C |
Accuracy
(@ 20°C/68°F) |
±1%
of reading
± (0.05 µS/cm or 1 digit, whichever greater) |
±1%
of reading
± (0.03 ppm or 1 digit, whichever greater) |
±1%
of reading |
±0.4°C |
| Typical
EMC Deviation |
±1%
of reading |
±1%
of reading |
±1%
of reading |
±0.1
°C |
| Calibration |
1
point with 6 memorized buffers:
84, 1413 ,5000,12880, 80000 or 111800 µS/cm |
- |
1
point
with 50-HI7037 buffer
(optional) |
2
points
at 0 and 50°C
(plus ±1°C adjustment) |
| Temp.
Compensation |
Automatic
or Manual from 0 to 60°C (can be disabled to measure actual
conductivity) |
| Temp.
Coefficient |
0.00
to 6.00 %/°C (for EC and TDS only) Default value is 1.90%/°C |
| TDS
Factor |
0.40
to 0.80 (default value is 0.50) |
| Probe |
50-HI76309
4-ring probe, K=1 nominal and built-in temperature sensor (included) |
| Power
Supply |
4x1.5V
AA alkaline type batteries (included), or 12 VDC adapter |
| Auto
Off |
After
5 minutes (can be disabled) |
| Casing |
IP
67 |
| Environment |
32
to 122°F (0 to 50°C); RH 100% |
| Dimensions |
7.7
x 3.1 x 2.4" (196 x 80 x 60 mm) |
| Weight |
18
oz (500 g) |
| (*)
Actual conductivity (or TDS) is the non-temperature compensated
conductivity (or TDS) value of a solution. |
ScienceLab offers Auto-Ranging ATC Conductivity/TDS/Temp/NaCl Meter and many other laboratory products and chemicals for all of your research project needs. Whether you work in a lab setting or perform testing in the field, we have the products you need, from Portable Conductivity/TDS Meters, glassware, chemicals, or testing equipment. We care about our customers' needs, if you are unable to find the scientific product your looking for then please contact us for assistance at 800-901-7247.
|
|