Hanna pH/Temp (pH210) & ph/mV/ORP (pH211) Bench Meters
The new 50-pH210 and 50-pH211 have
been designed to be simple without sacrificing accuracy. Both meters measure pH
from 0 to 14 and temperature from 0 to 100 with 0.01 pH and 0.5°C accuracy. 50-pH211
can also measure mV in two ranges of ±399.9 and ±1999 mV in conjunction with
ISEs and ORP electrodes. The large horizontal display shows both pH (or mV)
and temperature at the same time.
Calibration has been reduced to a few simple
steps, all guided by the meters sophisticated custom microprocessor. The user
can calibrate the meter at one or two points aided by automatic buffer
recognition at pH 4.01, 6.86, 7.01, 9.18 and 10.01. The display symbols also
guide the operator during the calibration procedure and provide a stability
indicator to ensure subsequent measurements are correct. In order to help you
make the most of your 50-pH210 and 50-pH211, the front label
carries clear and concise information about operation and calibration. This way
you do not have to take time during your busy schedule looking for instructions
manual.
All pH measurements are automatically compensated
for the temperature variance with the probe attached. Temperature can also be
manually compensated, if so desired. Simply press a key to memorize and later
recall the measurement. Both meters are supplied complete and ready-to-use with
a laboratory-grade glass pH electrode, stainless steel temperature probe, 12 VDC
transformer and even the calibration buffers.
50-pH210 is supplied complete with 50-HI1131B
interchangeable pH electrode, BNC, 1 m (3.3) cable, 50-HI7669/2W temp.
probe, pH 4 and 7 cal. sol., electrolyte sol., 12VDC adapter.
50-pH211 is supplied complete with 50-HI1131B
interchangeable pH electrode, BNC, 1 m (3.3) cable, 50-HI7669/2W temp.
probe, pH 4 and 7 cal. sol., electrolyte sol., 12VDC adapter.
| Specifications: |
50-pH210 |
50-pH211 |
| Range |
pH |
0.00
to 14.00 |
0.00
to 14.00 |
| mV |
- |
±399.9
mV (ISE); ±1999 mV (ORP) |
| °C |
0.0
to 100.0 |
0.0
to 100.0 |
| Resolution |
pH |
0.01 |
0.01 |
| mV |
- |
0.1
mV (ISE); 1 mV (ORP) |
| °C |
0.1 |
0.1 |
| Accuracy
(@68°F/20°C)
|
pH |
±0.01 |
±0.01 |
| mV |
- |
0.1
mV (ISE); 1 mV (ORP) |
| °C |
±0.5 |
±0.5 |
| Typical
EMC |
pH |
±0.03 |
±0.03 |
| Deviation |
mV |
- |
±2 |
| °C |
±0.3 |
±0.3 |
| pH
Calibration |
Automatic
1 or 2 points with 5 memorized buffers (pH 4.01, 6.86, 7.01, 9.18,
10.01) |
Temperature
Compensation |
Automatic
(with probe 50-HI7669/2W) or manual from 0 to 100°C (32 to 212°F) |
| Electrode |
50-HI1131B
pH electrode, combination, glass, interchangeable, BNC connector, 1 m
(3.3) cable (included) |
| Temperature
Probe |
50-HI7669/2W
with 1 m (3.3) cable (included) |
| Input
Impedance |
1012
ohm |
| Power
Supply |
12VDC
(included) |
| Environment |
32
to 122°F (0 to 50 °C); 95% R.H. |
| Dimensions |
240
x 182 x 74 mm (9.4 x 7.2 x 2.9) |
| Weight |
1.0
Kg (2.2 lb.) |
ScienceLab offers Hanna pH/Temp (pH210) & ph/mV/ORP (pH211) Bench Meters and many other laboratory products and chemicals for all of your research project needs. Whether you work in a lab setting or perform testing in the field, we have the products you need, from pH Meters - Bench Top, glassware, chemicals, or testing equipment. We care about our customers' needs, if you are unable to find the scientific product your looking for then please contact us for assistance at 800-901-7247.
|
|
Featured Products
|
|